| Model Number |
470309-12 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) has received the mega suturecut needle driver instrument involved with this complaint and completed the device evaluation.Failure analysis investigations found the instrument to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.The clevis did not exhibit any damage or wear marks.Other cables at the wrist were not damaged.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis suggests that if the malfunction were to recur it could cause or contribute to an adverse event.
|
| |
|
Event or Problem Description
|
|
The mega suturcut needle driver instrument was returned to intuitive surgical, inc.(isi) without a reported failure.Isi made multiple follow up attempts to obtain additional information, however, no information has been received as of date of this report.
|
| |
|