| Model Number |
VLOCL0024 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the device loop was broken, and the same event occurred repeatedly during the same surgery.The procedure was completed with another device.The surgical time was extended by less than 30 min.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary post market vigilance (pmv) led an evaluation of six devices.The visual inspection of the sample noted six sutures and six needles.One of the sutures was received with the broken loop; the loop wasn¿t returned.The rest of the sutures were received with the loops open; it was observed, under magnification, that each loop appeared to have split under tension.Upon microscopic inspection, evidence of material transfer was found at the weld site of each loop.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|