Model Number 42-7558
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 12/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was reported to be available but not returned.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 31-jan-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.(b)(4).
 
Event or Problem Description
It was reported that on (b)(6) 2017 the patient was scheduled to have the existing corflo radiologically placed nasojejunal tube (njt) advanced.The tip of the existing tube, which was placed (b)(6) 2017, had retracted into the duodenum.The main body of the tube had coiled into the patient's stomach after the patient had vomited.An initial x-ray noted the njt to be coiled in the stomach, so a 260 cm stiff terumo guide wire was inserted down the tube.When the wire was partially inserted the healthcare practioner proceeded to withdraw the njt under fluoroscopic guidance in order to uncoil the loop in the stomach in an effort to aid in advancement.Initially the looped tube was seen to move as the healthcare practioner withdrew the tube.The healthcare practioner continued to withdraw the tube.It became apparent that the tube had fractured into two pieces and the proximal portion was removed via the nose.Confirmation via x-ray noted the distal 55 cm of the njt to be in the patient¿s stomach.Additional information received 11-jan-2018 stated the healthcare practioner recovered all parts of the tube.The patient required an endoscopy procedure for removal of the retained device component.There was no reported patient injury.
 
Additional Manufacturer Narrative
One used sample was received.No product packaging was received.The product did not contain a lot number identification.The tubing exhibited discoloration from the point of separation to approximately the 75 cm marking.At the 54 cm mark, the tubing exhibited stretching or "ballooning" and had ruptured and separated.The detached distal portion of the tube was not returned for evaluation.Water was flushed through the feeding adapter on the proximal end of the tube.No occlusion or restriction of flow was observed.No other damage was observed on the sample.The root cause was inappropriate use.This issue was caused by excessive pressure caused by trying to unclog the tube by flushing.All information reasonably known as of 10-apr-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Additional Manufacturer Narrative
Additional information: sample device returned.All information reasonably known as of (b)(6) 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
HALYARD HEALTH
5405 windward parkway
alpharetta GA 30004
MDR Report Key7233695
Report Number3009124963-2018-00006
Device Sequence Number1656094
Product Code KNT
UDI-Device Identifier10815149020591
UDI-Public10815149020591
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other Health Care Professional
Device Model Number42-7558
Device Catalogue Number104608300
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2018
Initial Date Received by Manufacturer 01/08/2018
Supplement Date Received by Manufacturer02/08/2018
03/16/2018
Initial Report FDA Received Date02/01/2018
Supplement Report FDA Received Date02/20/2018
04/10/2018
Patient Sequence Number1
-
-