• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE: SYNTHES USA PRODUCTS LLC INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS; GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE: SYNTHES USA PRODUCTS LLC INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS; GUIDE Back to Search Results
Catalog Number 357.411
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information was not provided for reporting.Date of event is unknown.Device is an instrument and is not implanted/explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Device evaluation/investigation could not be completed; no conclusion could be drawn as product was not returned to manufacturer.Device history records review could not be completed without lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
(b)(6) reported as follow: it was reported that on an unknown date during the surgery, insertion handle for trochanteric fixation nail was bent or damaged and did not allow the aiming arm to align with the hole in the nail.There were no fragments generated, and there are no missing pieces.The surgery was completed using the device, with no reported patient harm.Concomitant device reported: tfn aiming arm (part # unknown, lot # unknown, quantity 1).This report is for one (1) insertion handle f/trochanteric fixation nails.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: lot number, udi.Date returned to manufacturer.Date of manufacture.Dhr review was completed.Device history record (dhr) review: part#: 357.411, lot#: 3328r16.Manufactured by supplier aeromed inc.Manufacturing date: 16-march-2010.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Customer quality conducted an investigation of the returned device.The 357.411 insertion handle is a reusable instrument routinely used in the titanium trochanteric fixation nail (tfn) system to aid in femoral nail insertion.The returned 8+ year old insertion handle was received at customer quality (cq) with significant post manufacturing damage.The device has several dents and scratches including hammer marks on the "do not strike" etch.Therefore this complaint is confirmed for damage.However the complaint for misalignment was not able to be confirmed or replicated via functional test because the other devices used in the construct were not returned.The damage in the holes could contribute to the reported complaint condition.A dimensional inspection of features relevant to this complaint could not be obtained at cq due to the post manufacturing damage.Drawings manufactured through current were reviewed during this investigation.No product design issues or discrepancies were observed.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The material and relevant material properties were determined to be conforming at the time of manufacture based on review of the dhr.A definitive root cause was not able to be determined.It is likely that the formation of dents and hammer marks were caused by misplaced hammer strikes and rough handling over the device's 8+ year service life which have led to this complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERTION HANDLE F/TROCHANTERIC FIXATION NAILS
Common Device Name
GUIDE
Manufacturer (Section D)
WRIGHTS LANE: SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7234158
Report Number2939274-2018-50467
Device Sequence Number1994764
Product Code FZX
UDI-Device Identifier10886982196347
UDI-Public(01)10886982196347(10)LOT#UNKNOWN
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number357.411
Device Lot Number3328R16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2018
Initial Date Received by Manufacturer 01/11/2018
Supplement Date Received by Manufacturer04/09/2018
Initial Report FDA Received Date02/01/2018
Supplement Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-