| Model Number |
TRUETRACK |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
Weakness (2145)
|
| Date of Event |
01/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Manufacturer contacted customer on (b)(6) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results accompanied by symptoms.Customer's wife is calling on his behalf.The customer is concerned with results obtained of 169,197,171 and 208 mg/dl.The expected fasting blood glucose test result range is 100-130 mg/dl.The customer did report symptoms of feeling weak.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the living room.During the call on (b)(6) 2018, a back to back blood test was not performed.The test strip lot manufacturer's expiration date is 04/26/2020 and open vial date is one month ago.The meter memory was reviewed for previous test result history: (b)(6).Customer states that he feels weak and does not know if that is related to his blood sugar.Customer states that he has been a diabetic for a long time and knows that his results should not be higher than 140 mg/dl fasting.Customer is concerned that the meter is giving high results when there have been no changes to his routine.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results accompanied by symptoms.Customer's wife is calling on his behalf.The customer is concerned with results obtained of 169,197,171 and 208mg/dl.The expected fasting blood glucose test result range is 100-130mg/dl.The customer did report symptoms of feeling weak.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the living room.During the call on (b)(6) 2018, a back to back blood test was not performed.The test strip lot manufacturer's expiration date is 04/26/2020 and open vial date is one month ago.The meter memory was reviewed for previous test result history: (b)(6).Customer states that he feels weak and does not know if that is related to his blood sugar.Customer states that he has been a diabetic for a long time and knows that his results should not be higher than 140mg/dl fasting.Customer is concerned that the meter is giving high results when there have been no changes to his routine.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Note: manufacturer contacted customer on 1/17/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|