• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABIOMED, INC. IMPELLA CP; TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABIOMED, INC. IMPELLA CP; TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP Back to Search Results
Model Number IMPELLA CP
Medical Device Problem Code Physical Property Issue (3008)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 01/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The products have not been returned from the user facility.(b)(4).
 
Event or Problem Description
An impella cp was implanted on (b)(6).There was difficulty inserting the repositioning sheath completely in, and the physician attempted to reposition it.The patient received two units of blood due to bleeding from the insertion site during the case.Upon removal of the impella cp, the repositioning sheath appeared to be damaged.
 
Additional Manufacturer Narrative
The impella cp was returned for investigation.The wound closure tip was found to be damaged with two tears noted.A closer inspection revealed there were visible scrape marks on the wound closure, leading directly into the cracks at the tip.The root cause of the access bleeding was the fractured wound closure tip.The root cause of the wound closure damage was unable to be determined.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPELLA CP
Common Device Name
TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP
Manufacturer (Section D)
ABIOMED, INC.
22 cherry hill drive
danvers MA 01923
MDR Report Key7234380
Report Number1220648-2018-00006
Device Sequence Number13867041
Product Code OZD
UDI-Device Identifier00813502011258
UDI-Public00813502011258
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P14000013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/30/2019
Device Model NumberIMPELLA CP
Device Catalogue Number0048-0003
Device Lot Number1311823
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/26/2018
Supplement Date Received by Manufacturer01/26/2018
Initial Report FDA Received Date02/01/2018
Supplement Report FDA Received Date02/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age43 YR
-
-