Model Number 728246 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint pr# (b)(4).
|
|
Event Description
|
Engineering analysis concluded that this event has the potential to result in image misinterpretation due to an artifact from a degraded compensator within the collimator.Therefore, this issue has been determined to be a reportable event.
|
|
Manufacturer Narrative
|
The customer reported a band artifact on patient images during head scans and an x-ray system can't comply issue (addressed in (b)(4)).There was no report of misrepresentation as a result of the band artifacts.The philips field service engineer (fse) evaluated the system and determined there was a crack in the compensator.The fse replaced the a-plane collimator and ran a full calibration to resolve the issue.The philips field service engineer (fse) confirmed there was no harm to a patient and no report of misrepresentation and/or mistreatment as a result of this reported issue.The system is in clinical use.This issue was determined not to be a reportable event.
|
|