• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE-R READY INDICATOR RED CELLS; ANTIGLOBULIN COATED RED CELLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE-R READY INDICATOR RED CELLS; ANTIGLOBULIN COATED RED CELLS Back to Search Results
Catalog Number 0006428
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The full number for the product in question is bk020053 ((b)(4) 2003).Immucor technical support used a remote elctronic connection method on (b)(4) 2018 to assess the instrument test well image in question, and determined that the instrument test well images in question were consistent with the instrument interpretation.
 
Event or Problem Description
On (b)(6) 2017, an immucor employee visiting a customer site reported an unexpected negative antibody screen when using capture-r ready indicator red cells on a galileo neo instrument, when tested on (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY INDICATOR RED CELLS
Common Device Name
ANTIGLOBULIN COATED RED CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key7235327
Report Number1034569-2018-00027
Device Sequence Number111576
Product Code KSF
UDI-Device Identifier10888234000600
UDI-Public10888234000600
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
BK020053
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Medical Technologist
Device Expiration Date01/21/2018
Device Catalogue Number0006428
Device Lot Number221010
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2018
Initial Report FDA Received Date02/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-