TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report # (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip: (b)(4).Note: manufacturer contacted customer on 1/19/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with results obtained results of 188, 169, 181 and 196mg/dl.The expected fasting blood glucose test result range is 100 to 150mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a blood test was performed fasting by the customer and produced test results of 171 mg/dl using true track meter.The test strip lot manufacturer's expiration date is 04/13/2020 and open vial date is a week ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).
 
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on 1/19/2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with results obtained results of 188, 169, 181 and 196 mg/dl.The expected fasting blood glucose test result range is 100 to 150 mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored by customer according to specification in the bedroom.During the call on (b)(6) 2018, a blood test was performed fasting by the customer and produced test results of 171 mg/dl using true track meter.The test strip lot manufacturer's expiration date is 04/13/2020 and open vial date is a week ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7235634
Report Number1000113657-2018-00130
Device Sequence Number111581
Product Code NBW
UDI-Device Identifier00021292008888
UDI-Public(01)00021292008888(17)200413(10)RU5177
Combination Product (Y/N)N
Initial Reporter StateUT
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRU5177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 01/11/2018
Supplement Date Received by Manufacturer01/11/2018
Initial Report FDA Received Date02/02/2018
Supplement Report FDA Received Date02/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY; SECOND THERAPY
-
-