• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 728306
Device Problems Signal Artifact/Noise (1036); Degraded (1153); Poor Quality Image (1408)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint (b)(4).
 
Event Description
Engineering analysis concluded that this event has the potential to result in image misinterpretation due to an artifact from a degraded compensator within the collimator.Therefore, this issue has been determined to be a reportable event.
 
Manufacturer Narrative
The customer reported there was a ring artifact present on images.The philips field service engineer (fse) went to the site to evaluate the system.The fse confirmed the reported issue that artifacts were present on scan images.The fse visually inspected the system and found a cracked compensator was the cause of the ring artifact.The fse replaced the compensator in the a-plane collimator to resolve the issue.The philips field service engineer (fse) confirmed there was no impact to the patient as a result of this issue.The system is in clinical use.This issue has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
MDR Report Key7235651
MDR Text Key99004526
Report Number1525965-2018-00188
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/21/2015
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-