Catalog Number 1125275-15
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was returned for analysis.The reported hub crack was able to be confirmed.A leak was noted coming from the crack on the hub.As the customer reported there was no leak during use of the device, the investigation determined the reported difficulties appear to be related to circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.There is no indication of a product quality issue with respect to design, manufacture or labeling.
 
Event or Problem Description
It was reported that the 2.75x15 mm xience alpine stent delivery system (sds) was advanced to the left anterior descending coronary artery target lesion.The hub of the sds cracked when it was being attached to the indeflator.The hub did not leak when it cracked.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.Device analysis identified a leak coming from the crack on the hub of the xience alpine sds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7235929
Report Number2024168-2018-00762
Device Sequence Number19087069
Product Code NIQ
UDI-Device Identifier08717648199257
UDI-Public08717648199257
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2020
Device Catalogue Number1125275-15
Device Lot Number7063041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2018
Initial Report FDA Received Date02/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INFLATION: INDEFLATOR
-
-