Supposed defective products were not returned to manufacturer so it was not possible to make an evaluation.The whole batch lot was composed by (b)(4) pcs, sold worldwide and no such claims have been received.During the quality test performed in production nothing was wrong.Possible explanation to the problem: the handle does not detach from the needle until any metal deformation occurs.Metal cannula used is a standard aisi 304 tube o.D.3.0 mm, i.D.2.3 mm, if the medical device is used inside a very hard bone it could happen that the needle cannula can steaks inside the bone, under these conditions following the ifu, the device must be removed from the bone rotating the handle one turn on the right and one turn on the left in order to avoid any metal cannula deformation and after that the plastic handle detachment.The fact that two needles had the same problem maybe confirms that this detachment occurred in these conditions and that operator used the needle not in the right way as described in the ifu.We already informed our distributor about this problem in december 2017.(b)(4).
|