| Catalog Number |
02.12.0312FL |
| Health Effect - Clinical Code |
Joint Disorder (2373)
|
| Date of Event |
09/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Clinical evaluation performed by medical affairs manager on oct 5, 2018: insert revision in tka 10 months after primary for joint instability.The insert was exchanged to a thicker one.The developmental instability is a rather commonly described possible complication following tka, because soft tissue may stretch and provide insufficient stability.It is therefore common practice to resort to a thicker insert and recreate soft tissue tension.This is not a problem due to any implant defect or malfunction.Batch review performed on 16 october 2018: lot 170292: (b)(4) items manufactured and released on 26 april 2017.Expiration date: 2022-04-06.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
Revision surgery performed due to knee instability after about 1 year from primary.The surgeon revised the 12mm liner to a 14mm liner.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2018 we were informed that the revision surgery has been successfully performed on (b)(6) 2018.
|
| |
|
Event or Problem Description
|
|
The surgeon revised the patient 8 months after primary for a knee instability.
|
| |
|