MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Halo (2227); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2016 and (b)(6) 2017.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zlb00 24.0 diopter intraocular lens (iol) was implanted in the patient's left eye (os) on (b)(6) 2016.It was later explanted on (b)(6) 2017 due to the patient complaining of blurry vision and severe halo/glare.The lens was replaced with a zcb00 25.0 diopter iol.Reportedly, the patient's post-operative end refraction was 20/30 -0.25 x -.025 x 157 20/20.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search in the complaints history revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7237866
Report Number9614546-2018-00107
Device Sequence Number13661215
Product Code MFK
UDI-Device Identifier05050474560758
UDI-Public(01)05050474560758(17)190623
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/08/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date06/23/2019
Device Model NumberZLB00
Device Catalogue NumberZLB00U0240
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/03/2018
Supplement Date Received by Manufacturer07/31/2018
Initial Report FDA Received Date02/02/2018
Supplement Report FDA Received Date08/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Date Report Sent to FDA01/01/2005
-
-