MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY WALLFLEX¿ BILIARY RX STENT SYSTEM; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY WALLFLEX¿ BILIARY RX STENT SYSTEM; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00570660
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device has been disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
Note: this manufacturer report pertains to one of two devices used in the same procedure.Refer to manufacturer report # 3005099803-2018-00144 and this report for the associated device information.It was reported to boston scientific corporation on (b)(6) 2018 that two wallflex biliary rx uncovered stents were to be used to treat an approximately 6-7 cm malignant lesion in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2018.Reportedly, the patient's anatomy was very tight and was not dilated prior to stent placement.According to the complainant, during the procedure, the physician was able to deploy the first stent (the subject of this report) and the second wallflex biliary rx stent (subject of this report).However, a couple of the loops on the end of the two stents were broken into pieces.The physician removed both stents using forceps and the procedure was completed with a different device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX¿ BILIARY RX STENT SYSTEM
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7238014
Report Number3005099803-2018-00145
Device Sequence Number13719654
Product Code FGE
UDI-Device Identifier08714729759287
UDI-Public08714729759287
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K140630
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2019
Device Model NumberM00570660
Device Catalogue Number7066
Device Lot Number0020776890
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2018
Initial Report FDA Received Date02/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-