Catalog Number ACRO-25-450
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: our evaluation of the product said to be involved confirmed the report.The core wire of the device separated at 7.2 cm with the wire guide covering surrounding the break fully intact.The wire guide covering at the core wire break is slightly bent, but is not otherwise stretched or deformed.The wire guide is kinked approximately 15.1 cm from the distal end.A rough surface was felt on the wire guide covering from approximately 30 cm to 35 cm.During evaluation, the wire guide covering was separated and the cross section was evaluated under magnification.Both ends of the broken core wire are slightly bent with no noticeable jagged edges.The diameter at both ends of the broken core wire were measured and confirmed to be the same (a bottle neck in the taper/grind was not found).A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Prior to distribution, all cook acrobat calibrated tip wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an isolated occurrence.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a gastroscopy, the physician used a cook acrobat calibrated tip wire guide.The wire was broken into two (2) parts when withdrawn.The procedure was completed with another manufacturer's guidewire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT¿ CALIBRATED TIP WIRE GUIDE
Common Device Name
OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key7238021
Report Number1037905-2018-00040
Device Sequence Number16447414
Product Code OCY
UDI-Device Identifier00827002327597
UDI-Public(01)00827002327597(17)200209(10)W3826990
Combination Product (Y/N)N
Initial Reporter CountryTW
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberACRO-25-450
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date01/08/2018
Device Age11 MO
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2018
Initial Report FDA Received Date02/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
Patient Weight78
-
-