MAUDE Adverse Event Report: BAYER MEDICAL CARE, INC. SPECTRIS SOLARIS SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYER MEDICAL CARE, INC. SPECTRIS SOLARIS SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT Back to Search Results
Model Number 60728605
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the limited information, we are unable to determine the root cause of the unknown particle at this time; however, the product is scheduled to return to bayer for a full evaluation.Once the evaluation is completed, a follow-up report will be submitted.
 
Event or Problem Description
The customer reported the following: after opening the solaris mr disposable kit and upon inspection, they saw a foreign body within the low pressure connector tubing (lpct).The customer elected not to use the disposable kit.No injury or adverse event was reported.
 
Additional Manufacturer Narrative
Bayer product analysis received and examined one returned low pressure connector tubing (lpct).Visual examination confirmed the presence of a particle within the tube set.The particle was identified as degraded resin from the tubing extrusion process and was partially connected to the inside wall of the tubing.As the particle was partially embedded into the tubing wall, there is a potential that it could fragment while under pressures typically generated during an injection.Therefore, the potential of injection into a patient exists.Our investigation determined that the particle noted in the lpct was introduced during the manufacturing process and was not detected upon receipt of the product from the supplier.A 12 month review of complaint history determined the complaint rate to be 0.57 complaints per million (cpm).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRIS SOLARIS SYRINGE KIT
Common Device Name
ANGIOGRAPHIC SYRINGE KIT
Manufacturer (Section D)
BAYER MEDICAL CARE, INC.
1 bayer drive
indianola PA 15051
MDR Report Key7238038
Report Number2520313-2018-00020
Device Sequence Number1222666
Product Code DXT
UDI-Device Identifier00616258005461
UDI-Public(01)00616258005461
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K012950
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 02/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2022
Device Model Number60728605
Device Catalogue NumberSSQK 115
Device Lot Number8405966
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2018
Initial Date Received by Manufacturer 01/16/2018
Supplement Date Received by Manufacturer01/16/2018
Initial Report FDA Received Date02/02/2018
Supplement Report FDA Received Date02/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-