LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-50
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 12/06/2016
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-50 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).This event was initially considered to be non-reportable.However, after additional evaluation, livanova (b)(4) has decided to reclassify this event as reportable.This report is being filed as part of a retrospective review conducted in response to this new decision.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was sent to the facility to investigate.The field representative was able to confirm the issue and traced it back to a faulty compressor.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Evaluated on site by livanova technician.
 
Event Description
Livanova (b)(4) received a report that the heater-cooler system 3t blew the fuses when switched on during setup.There was no patient involvement.
 
Manufacturer Narrative
During complaint investigation, it was determined that the reported issue is not related to a heater-cooler system 3t.The reported issue is related to a heater-cooler system 1t, which is not marketed in the united states.As the involved device is not marketed in the united states, the event has been reassessed as non-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7238080
MDR Text Key99102486
Report Number9611109-2018-00084
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-50
Date Manufacturer Received03/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-