• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL, INC STATIC AND DYNAMIC COMPLIANCE; CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC STATIC AND DYNAMIC COMPLIANCE; CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION Back to Search Results
Model Number VMAX ENCORE 22
Device Problem Calibration Problem (2890)
Patient Problem No Patient Involvement (2645)
Event Date 05/18/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The suspected device was received and a failure investigation was performed.The reported issue was confirmed and a burnt analyzer board during was found.Upon further investigation, the failure analysis technician found that the hex nut went through the tracing causing a short, due to a washer not being installed.
 
Event Description
The customer reported that the vmax module is having intermittent issues with lung diffusion (dlco) testing.There was no patient involvement at the time of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATIC AND DYNAMIC COMPLIANCE
Type of Device
CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7238567
MDR Text Key98997497
Report Number2021710-2018-07374
Device Sequence Number1
Product Code BTY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K981366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberVMAX ENCORE 22
Device Catalogue Number777404-101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-