MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551NAP
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 01/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that the insulin pump was not vibrating.Customer's blood glucose value was 270mg/dl.The last couple of days where its been 300mg/dl, 250mg/dl ,275mg/dl.Customer treated with the pump.Customer also believed that his pump is not delivering basals.Insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with intermittent button response due to flattened (escape, up and act ) button dome switch (no crease) and partial unlocked lcd keypad connector noted during visual inspect.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime and excessive no delivery test or verify audio or vibrate anomaly and possible under delivery anomaly due to buttons not responding.Device received with cracked battery tube threads, cracked lcd window, minor scratched lcd window, cracked reservoir tube lip, cracked case display window corner, stained end cap sticker, stained address or serial number label, cracked belt clip slot and cracked case reservoir tube window corner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7238671
Report Number3004209178-2018-48059
Device Sequence Number1146098
Product Code OZO
UDI-Device Identifier00643169736047
UDI-Public(01)00643169736047
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-551NAP
Device Catalogue NumberMMT-551NAP
Device Lot NumberA6551NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2018
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer05/18/2018
Initial Report FDA Received Date02/02/2018
Supplement Report FDA Received Date06/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age29 YR
Patient Weight215
-
-