Model Number 8637-40
Device Problems Kinked (1339); Aspiration Issue (2883); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2017
Event Type  Injury  
Manufacturer Narrative
The main component of the system and other applicable components are: product id: 8780, serial# (b)(4), implanted: (b)(6) 2016, product type: catheter.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacture representative (rep) via a healthcare provider (hcp) regarding a patient receiving unknown drug via an implanted pump.It was reported the patient had a catheter revision and pump replacement on (b)(6) 2017.It was noted the cause of the pump and catheter replacement was that they tried to do a dye study, but were unable to push anything through the catheter.It was mentioned a catheter kink was found during revision which had caused back pressure to the pump and resulted in the pump as well as the catheter being replaced.The issue was resolved once the pump and catheter were replaced.The rep noted they still had the pump and catheter in their procession and was going to be sending it back.The patient's weight was not known.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7238697
MDR Text Key98961959
Report Number3004209178-2018-02007
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169508156
UDI-Public00643169508156
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2018
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/25/2018
Date Device Manufactured09/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age31 YR
-
-