Model Number MMT-523LNAS
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer was reported via phone call that, the customer had high blood glucose.Customer did not believe her pump was giving her the current amount of insulin.Customer stated that, after she does a bolus her blood glucose is still rising.The current blood glucose was 252 mg/dl.Customer treated high blood glucose with bolus.Based on customer report proceeding in troubleshooting per customer alleging possible pump under delivery.The drive support cap appears normal.Run manual prime and fill tubing with current set and insulin exited at the quick release set.Customer advised to change entire set, reservoir and insulin and treat per healthcare professional instructions.The device will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7239228
Report Number3004209178-2018-48129
Device Sequence Number16449790
Product Code OYC
UDI-Device Identifier00643169479630
UDI-Public(01)00643169479630
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Device Lot NumberA3523LNASJ
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2018
Initial Report FDA Received Date02/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
Patient Weight168
-
-