MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. 16" CENTURY STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. 16" CENTURY STERILIZER Back to Search Results
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/05/2018
Event Type  malfunction  
Manufacturer Narrative
No report of injury, procedure delay or cancellation. A steris service technician inspected the sterilizer and could not detect an odor; no adverse effects from the odor were reported. The technician found that the safety valve on the steam generator was damaged subsequently allowing steam to leak. The technician replaced the safety valve, ran a test cycle and confirmed the unit to be operating according to specification. The sterilizer is not under steris service agreement; all preventive maintenance is performed by the user facility. The unit was installed at the customer's location in 2009, and no additional issues have been reported.
 
Event Description
The user facility reported that steam was leaking from their 16" century sterilizer and an odor was emitting from the unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name16" CENTURY STERILIZER
Type of DeviceSTERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key7239304
MDR Text Key99101527
Report Number3005899764-2018-00005
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/02/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/05/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-