A user facility nurse reported that the venous chamber would not hold prime at the start of a patient¿s hemodialysis (hd) treatment, allowing air to enter the bloodlines.Prior to the start of hd treatment, the fluid in the venous chamber was dropping, so the nurse filled the lines with saline during priming.When the patient¿s blood was in the venous chamber, it was reported that the blood dropped so low, air had entered the bloodlines.There was no machine alarm.The nurse removed the venous bloodline section and replaced it with another venous line from the same lot number.The nurse stated that the same thing happened where the blood dropped below level, allowing air to enter in the bloodline.The patient¿s blood was not returned from both sets of venous bloodlines and the total estimated blood loss (ebl) was noted as more than 200ml.The patient completed treatment with a set-up of new supplies on a different machine.No patient adverse effects or injuries were experienced, and no medical intervention was required.The patient was able to complete treatment without any further interruptions or complications.The patient¿s hemoglobin was tested after treatment as the patient had a low hemoglobin level prior to the start of treatment, however, no actions were taken regarding the blood loss.The patient continues regularly scheduled hd treatments without issue.The machine was removed from service following the event.The user facility biomedical engineer (biomed) evaluated the machine and confirmed it was functioning properly.The machine was still sequestered pending clinical approval prior to putting back into service.There was no reported defect or crack with either set of bloodlines used.The bloodline device was stated to be available to be returned to the manufacturer for evaluation.
|