Catalog Number 157012165
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 01/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Der states patient fell while skiing resulting in a periprosthetic femur fracture around and distal to summit stem.Surgeon was able to explant the size 9hi summit stem and 36+1.5 ceramic revision femoral head without any complications or bone loss.Surgeon used a stryker restoration modular stem and cables to fix the patients fracture.The pinnacle 58 cub and 58+4n liner was well fixed.It was also indicated that the stem was loose due to fracture involving bone around femoral stem.Doi: (b)(6) 2016; dor: (b)(6) 2018; left hip.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ9 HI OFF
Common Device Name
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key7240640
Report Number1818910-2018-52707
Device Sequence Number19136507
Product Code LPH
UDI-Device Identifier10603295060208
UDI-Public10603295060208
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K011489
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number157012165
Device Lot Number382242
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/17/2018
Supplement Date Received by Manufacturer03/01/2018
Initial Report FDA Received Date02/05/2018
Supplement Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-