• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C2860
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578); Device Handling Problem (3265)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2018-00233; 3005168196-2018-00234; 3005168196-2018-00235; 3005168196-2018-00236.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a pulmonary arteriovenous malformation (pavm) using ruby coils.It was noted that the patient's anatomy was tortuous.During the procedure, while attempting to implant a ruby coil into the pavm using a lantern delivery microcatheter (lantern), the physician experienced resistance after advancing the ruby coil approximately 75% into the pavm and the coil would not advance further; therefore, it was removed.While attempting to advance a new ruby coil through the lantern, the physician experienced resistance; therefore, the lantern and coil were removed and the physician flushed clot out of the lantern.It was reported that the ruby coil was then accidentally dropped on the floor after removal from the lantern.The physician wanted to start fresh and decided to use a second lantern.After multiple attempts, the physician was able to get the lantern back engaged into the pavm and attempted to advance a new ruby coil through the lantern; however, resistance was encountered and the ruby coil would not advance.The physician then decided to end the procedure at this point and the patient will be scheduled for another procedure.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7240799
Report Number3005168196-2018-00232
Device Sequence Number10953613
Product Code HCG
UDI-Device Identifier00814548013091
UDI-Public00814548013091
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/14/2025
Device Catalogue NumberRBY2C2860
Device Lot NumberF77731
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2018
Initial Report FDA Received Date02/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age34 YR
-
-