Catalog Number 5MAXACE068KIT
Medical Device Problem Codes Kinked (1339); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the ace68 was kinked approximately 59.0 cm from the proximal end.Conclusions: evaluation of the returned device confirmed the ace68 was kinked.If the ace68 is withdrawn from the packaging prior to removing the tubing tray from the kit retainer, damage such as this may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a thrombectomy procedure, the hospital staff noticed that the penumbra system ace 68 reperfusion catheter (ace68) was kinked upon removal from the packaging of the penumbra system ace 68 hi-flow kit (kit).The damaged ace68 was found prior to use and therefore; it was not used in the procedure.The procedure was completed using a new ace68.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ACE 68 HI-FLOW KIT
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7240823
Report Number3005168196-2018-00241
Device Sequence Number1223301
Product Code NRY
UDI-Device Identifier00814548016597
UDI-Public00814548016597
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K161640
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/09/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/31/2019
Device Catalogue Number5MAXACE068KIT
Device Lot NumberF72344
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/09/2018
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date02/05/2018
Supplement Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-