| Catalog Number |
5MAXACE068KIT |
| Medical Device Problem Codes |
Kinked (1339); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the ace68 was kinked approximately 59.0 cm from the proximal end.Conclusions: evaluation of the returned device confirmed the ace68 was kinked.If the ace68 is withdrawn from the packaging prior to removing the tubing tray from the kit retainer, damage such as this may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
During preparation for a thrombectomy procedure, the hospital staff noticed that the penumbra system ace 68 reperfusion catheter (ace68) was kinked upon removal from the packaging of the penumbra system ace 68 hi-flow kit (kit).The damaged ace68 was found prior to use and therefore; it was not used in the procedure.The procedure was completed using a new ace68.
|
| |
|