PENUMBRA, INC. INDIGO SYSTEM CAT5 ASPIRATION CATHETER; DXE Back to Search Results
Catalog Number CAT5
Medical Device Problem Codes Break (1069); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the cat5 was fractured approximately 107.0 cm from the hub.Conclusions: evaluation of the returned device confirmed it was fractured.This damage may have occurred due to forceful withdrawal of the cat5 from the packaging at extreme angles.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a thrombectomy procedure using an indigo system cat5 aspiration catheter (cat5), the scrub technician removed the cat5 from the sterile package and noticed that the catheter was broken.The damage to the cat5 was found prior to use; therefore it was not used in the procedure.The procedure was completed using a new cat5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM CAT5 ASPIRATION CATHETER
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7240827
Report Number3005168196-2018-00238
Device Sequence Number11021936
Product Code DXE
UDI-Device Identifier00814548016092
UDI-Public00814548016092
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Radiologic Technologist
Device Expiration Date02/16/2020
Device Catalogue NumberCAT5
Device Lot NumberF74471
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2018
Initial Report FDA Received Date02/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-