Catalog Number 10492730
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens evaluated the data received.The investigation shows that the value obtained (167) from system 35109 was compromised by an air bubble in the sample.The air bubble covered part of the active electrode so the sensor could not properly reach endpoint and therefore reported a lower value.A plot of the patient's glucose recovery shows that the patient's glucose levels were increasing.The customer stated that there was no harm to the patient and no treatment or medical procedure was delayed or withheld.The measurement cartridge was replaced and the qc's were in the expected ranges.
 
Event or Problem Description
The customer reported discrepant glucose values on two different rp 500 analyzers.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Common Device Name
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
MDR Report Key7241642
Report Number3002637618-2018-00014
Device Sequence Number1016129
Product Code CHL
UDI-Device Identifier00630414589169
UDI-Public00630414589169
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K113216
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number10492730
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/12/2018
Initial Report FDA Received Date02/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-