| Model Number |
LF4318 |
| Medical Device Problem Codes |
Energy Output To Patient Tissue Incorrect (1209); Difficult or Delayed Activation (2577); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the device had an activation issue, wherein the handle was difficult to close to activate sealing.Follow up with the reporter provided additional information.The reporter stated that the handle and the activation button for the sealing would become stuck sometimes and handle closure was not smooth.The report also indicated that sealing was incomplete at times when activating the device, sometimes requiring several activations to achieve hemostasis.There was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device sample was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product met specification as received.Visual inspection found no defects.The reported conditions were not confirmed.The jaws opened and closed properly when the handle was retracted and released.The knife blade advanced and retracted smoothly along the knife track.The knife cut of the device was tested on a silicone test strip with acceptable results.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the device had an activation issue, wherein the handle was difficult to close to activate sealing.There was no patient injury.Reporter stated the handle and the activation button for the sealing became stuck sometimes.The closing of the handle was not smooth.Sealing was incomplete at times when device undergoing activation, several activations were required for hemostasis.
|
| |
|