Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer initially reported complaint for e-3 error: used test strip, test strip outside of vial too long, sample on top of test strip.The e-3 error had occurred once the test strip was inserted into the meter.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of hi and hi using truetrack meter.The customer did not know her expected blood glucose test result range.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer had just purchased the test strips on day of call; customer was informed of proper storage locations.The test strip lot manufacturer's expiration date is 4/13/2020 and open vial date is (b)(6) 2018.Customer was only able to obtain two results from the meter memory: result 1 : hi mg/dl date: (b)(6) 2018 time:07:09 pm fasting, result 2 : hi mg/dl date: (b)(6) 2018 time:07:09pm fasting.Customer called in states the meter is reading e-3 while conducting troubleshooting customer received an hi display.Customer states the meter read hi fasting and another hi fasting within 15 minutes of each other.Customer states she does not know her normal fasting range is.Customer could only obtain two readings of hi in history of meter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7242176
Report Number1000113657-2018-00144
Device Sequence Number13749264
Product Code NBW
UDI-Device Identifier00011822577540
UDI-Public(01)00011822577540
Combination Product (Y/N)N
Initial Reporter StateMI
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 02/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/13/2020
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRU5177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2018
Initial Report FDA Received Date02/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-