| Catalog Number |
329420 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported with the use of the bd micro-fine¿ iv insulin syringe the needle tips were breaking.No report of additional information.There was no report of injury or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Unable to perform complaint lot history check due to unknown lot number.Severity: s1; occurrence: unable to perform complaint lot history check due to unknown lot number.Based on the samples/photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Based on the above, no additional investigation and no capa is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.
|
| |
|