| Catalog Number |
PNS6F125SIM |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a medical procedure using a neuron 6f select catheter (6f select).During the procedure, the physician was unable to advance the 6f select into a neuron max 6f 088 long sheath (neuron max).It was reported that the 6f select did not fit into the neuron max.Therefore, the procedure was completed using another 6f select and the same neuron max.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no visible damage to the neuron 6f select catheter (6f select).The outer diameter (od) of the 6f select was measured using a ring gauge and found to be within specification.Evaluation of the returned device revealed that the 6f select od was within specification.The od of the 6f select was measured with a ring gauge and found to be within specification.Further evaluation of the returned device revealed that the 6f select was able to be advanced through a demonstration neuron max 6f 088 long sheath (neuron max) and out of the distal tip without an issue.The root cause of the reported issue that the 6f select was not fitting into the neuron max could not be determined.The neuron max referred to in the complaint was not returned to penumbra for evaluation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Search Alerts/Recalls
|