Model Number MMT-723LNAB
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call for unexpected restart.Customer¿s blood glucose was unknown.Customer reported that this happened starting with the last three batteries and they last more than a week.Customer reported that alarm occurred shortly after inserting a new battery.Customer reported that this issue occurs every time he changes his battery.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Pump passed all functional testing including the displacement, operating currents, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No alarm noted during test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7243179
Report Number3004209178-2018-48451
Device Sequence Number1043872
Product Code OYC
UDI-Device Identifier00643169478565
UDI-Public(01)00643169478565
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAB
Device Catalogue NumberMMT-723LNAB
Device Lot NumberA4723LNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2018
Initial Date Received by Manufacturer 01/25/2018
Supplement Date Received by Manufacturer06/27/2018
Initial Report FDA Received Date02/05/2018
Supplement Report FDA Received Date07/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR