Model Number ZCT400
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens remains implanted and therefore not explanted.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a toric intraocular lens (iol) was implanted in the right eye of a (b)(6) male patient on (b)(6) 2017.There were no patient complication and/or related injury.The lens originally placed at 118 degrees, was measured at 130 degrees on (b)(6) 2018.Reportedly, there are no plans for iol repositioning.The lens remains implanted in the eye.The patient was reported doing well.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site as to date it remains implanted.Therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7243197
Report Number9614546-2018-00116
Device Sequence Number882569
Product Code HQL
UDI-Device Identifier05050474553767
UDI-Public(01)05050474553767(17)180607
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source health professional,study,use
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/07/2018
Device Model NumberZCT400
Device Catalogue NumberZCT400U125
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/10/2018
Supplement Date Received by Manufacturer03/12/2018
Initial Report FDA Received Date02/05/2018
Supplement Report FDA Received Date04/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-