Troubleshooting was conducted with the customer, including inspecting the faceplate and back plate for damage; inspecting and cleaning the diaphragm; and disassembling, inspecting, cleaning and reassembling the kit and tubing.The troubleshooting did not resolve the customer's suction issue.The customer was sent a replacement pump, along with pump and save bags and hydrogel pads.Attempts to contact the customer to get additional information have gone unanswered.Based on the results of ca11-001, it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is 0.008% for the period of january 2013 to august 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history or mastitis." riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.Mastitis requires prompt medical attention for the mother for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
|