| Catalog Number |
313D |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
Infarction, Cerebral (1771)
|
| Date of Event |
01/04/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Information regarding patient age, weight and medical history were not provided.(b)(6).The product is not available for evaluation and testing as it remains implanted in the patient.Additional information will be submitted within 30 days upon receipt.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Event or Problem Description
|
|
As reported by a healthcare professional, the patient experienced an ischemic stroke immediately following an elective coil embolization of a 6 mm with 5 mm neck basilar tip aneurysm with a 3.5-4.5 mm 10 mm y-shape pulserider aneurysm neck reconstruction device (anrd) (313d/ w2590-12).Post-stroke neurological assessment revealed homonymous hemianopsia.Brain magnetic resonance imaging (mri) performed two days after the procedure revealed posterior cerebral artery (pca) ischemia with multiple scattered foci.The stroke was ¿expectantly¿ managed and the patient was discharged with residual stroke symptoms.The physician stated that due to the angle the ¿left arm [was] deployed in the aneurysm, not the pca¿.The geometric configuration of the aneurysm presented a challenge to the physician.One of the pcas was pointing superiorly 30 degrees while the left pca was pointing inferiorly.Due to the pcas¿ angulation, one leaflet was placed inside of the aneurysm and the other was placed inside of the aneurysm dome.The physician felt it was difficult to project how effectively the leaflet in the pca was protecting the origin of the artery and perhaps some of the coils that were placed during the procedure could have partially occluded the pca.Manufacturer sales representative was present during the case.The physician was instructed that the best position of the product should be in the pcas.Nonetheless, the product was selected was a y-configuration.There was no difficulty in trackability or stability noted during advancement of the product and deployment.The surgeon was pleased following the procedure as the product behaved as expected.Post-procedure magnetic resonance angiography (mra) demonstrated that both pcas were patent and there were no changes in flow.There was no evidence of a vessel or aneurysm rupture.The procedure was performed under heparinization and the patient was loaded with dual antiplatelet therapy (dapt).Stryker coils were utilized for the case.Of note, the physician has placed one pulserider prior to this case.The product remains implanted in the patient and is thus not available for return.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Complaint conclusion: as reported by a healthcare professional, the patient experienced an ischemic stroke immediately following an elective coil embolization of a 6mm with 5mm neck basilar tip aneurysm with a 3.5-4.5mm 10mm y-shape pulserider aneurysm neck reconstruction device (anrd) (313d/ w2590-12).Post-stroke neurological assessment revealed homonymous hemianopsia.Brain magnetic resonance imaging (mri) performed two days after the procedure revealed posterior cerebral artery (pca) ischemia with multiple scattered foci.The stroke was ¿expectantly¿ managed and the patient was discharged with residual stroke symptoms.The physician stated that due to the angle the ¿left arm [was] deployed in the aneurysm, not the pca¿.The geometric configuration of the aneurysm presented a challenge to the physician.One of the pcas was pointing superiorly 30 degrees while the left pca was pointing inferiorly.Due to the pcas¿ angulation, one leaflet was placed inside of the aneurysm and the other was placed inside of the aneurysm dome.The physician felt it was difficult to project how effectively the leaflet in the pca was protecting the origin of the artery and perhaps some of the coils that were placed during the procedure could have partially occluded the pca.Manufacturer sales representative was present during the case.There was no difficulty in trackability or stability noted during advancement of the product and deployment.The surgeon was pleased following the procedure as the product behaved as expected.Post-procedure magnetic resonance angiography (mra) demonstrated that both pcas were patent and there were no changes in flow.There was no evidence of a vessel or aneurysm rupture.The procedure was performed under heparinization and the patient was loaded with dual antiplatelet therapy (dapt).Stryker coils were utilized for the case.Of note, the physician has placed one pulserider prior to this case.Multiple attempts to obtain images were unsuccessful.No further information was provided.The product remains implanted in the patient and is thus not available for evaluation.A review of manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection processes related to the reported complaint.Stroke is a known potential complication with use of the pulserider.With the information available and without films of the event it is not possible to draw a clinical conclusion between the device and the reported event; however, the selection of the device configuration for the geometric characteristic as described by the physician was not ideal.Clinical and procedural factors including patient anatomy, vessel characteristics, device selection, and device manipulation are all factors that may contributed to the reported event.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Search Alerts/Recalls
|
|
|