C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETERIZATION TRAY WITH RED RUBBER CATHETER Back to Search Results
Model Number 774415
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter was missing the drainage eyes, which prevented the urine from draining.The device was replaced and no medical intervention was required.
 
Event or Problem Description
It was reported that the catheter was missing the drainage eyes, which prevented the urine from draining.The device was replaced and no medical intervention was required.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.Received 1 red urethral catheter with the original packaging for evaluation.The exterior of the sample was visually examined and found that there were no drainage eyes.A potential root cause is this may occur during the eye punching operation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box." (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URETHRAL CATHETERIZATION TRAY WITH RED RUBBER CATHETER
Common Device Name
CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7244331
Report Number1018233-2018-00357
Device Sequence Number1342831
Product Code EYB
UDI-Device Identifier00801741040221
UDI-Public(01)00801741040221
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model Number774415
Device Catalogue Number774415
Device Lot NumberNGBW4021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2018
Initial Date Received by Manufacturer 01/17/2018
Supplement Date Received by Manufacturer03/20/2018
Initial Report FDA Received Date02/06/2018
Supplement Report FDA Received Date03/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-