Model Number 304-20
Device Problem Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2017
Event Type  malfunction  
Event Description
A company representative who attended a patient¿s vns replacement surgery provided that when she connected the new generator to the lead systems diagnostics showed low impedance.A second diagnostics test also showed low impedance.The test resistor was connected to the generator and the impedance was ok.Pre-operative diagnostics were unable to be obtained as she was unable to interrogate the previous device.The physician then connected the lead to the new generator and performed system diagnostics two more times and the impedance was ok, both times.The lead was not replaced.The explanted device was discarded by the hospital.Additional relevant information has not been received to-date.No known surgery has occurred to-date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7244353
MDR Text Key99197357
Report Number1644487-2018-00172
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/20/2016
Device Model Number304-20
Device Lot Number3359
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 01/17/2018
Initial Date FDA Received02/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age33 YR
-
-