• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 38

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 38 Back to Search Results
Model Number 520-01-238
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Date of Event 01/31/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the glenoid being completely loose.The stem was still well fixed and in working condition.The surgeon removed the glenoid and installed the baseplate with glenosphere.He then removed the turon head, kept the stem and used reverse shoulder prosthesis (rsp) conversion with turon stem.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a loose glenoid.The previous surgery and the revision detailed in this investigation occurred 4 years and 9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a loose glenoid baseplate.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's loosening.There are many factors that may contribute to the event that are outside the control of djo surgical such as: loose joints from inadequate soft tissue support, degenerative bone, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PEGGED GLENOID IMPLANT, SIZE 38
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7245408
Report Number1644408-2018-00107
Device Sequence Number1028105
Product Code KWS
UDI-Device Identifier00888912024945
UDI-Public(01)00888912024945
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Model Number520-01-238
Device Catalogue Number520-01-238
Device Lot Number198G1025
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/31/2018
Supplement Date Received by Manufacturer03/15/2018
Initial Report FDA Received Date02/06/2018
Supplement Report FDA Received Date03/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000, LOT 878C1184; 520-42-116, LOT 913C1019; 520-00-000, LOT 878C1184; 520-42-116, LOT 913C1019
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-