• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS DOT PROV 4F DL FULL 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS DOT PROV 4F DL FULL 3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Catalog Number S1274108
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of rebu1047 showed one other similar product complaint from this lot number.
 
Event or Problem Description
The facility reported the catheter was replaced due to break/leak at the hub.No patient injury reported.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leak is confirmed.One 4 fr dual lumen powerpicc provena was returned for evaluation.An initial visual observation showed a large amount of blood residue throughout the sample, and the ink on the extension legs, the bifurcation, and the depth markers between 0 cm and 3 cm were observed worn and faded.A large split was observed in the extension tubing of the purple lumen just within the distal end of the luer connector.An attempt was made to flush both lumens and a spraying leak was observed emanating from the split in the purple lumen; however both lumens were not patent to infusion due to the large amount of blood residue within the sample.A microscopic observation revealed the fracture surface of the split in the extension tubing of the purple lumen was rough and exhibited cracks/fissures and beach marks, which are indicative of material fatigue.A small split was observed in the red lumen and had similar fracture surface characteristics as the split in the purple lumen.The tubing material of the both the purple and red lumens were observed to be attached to the luer connector at irregular points along the tubing¿s circumference, and cracking appeared to have initiated at one or more of these points in both lumens.A lot history review (lhr) of rebu1047 showed one other similar product complaint from this lot number.
 
Event or Problem Description
The facility reported the catheter was replaced due to break/leak at the hub.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROV 4F DL FULL 3CG
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7246259
Report Number3006260740-2018-00131
Device Sequence Number19427626
Product Code LJS
UDI-Device Identifier00801741130120
UDI-Public(01)00801741130120
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Health Professional
Remedial Action Recall,Other,Notification
Type of Report Initial,Followup
Report Date (Section B) 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberS1274108
Device Lot NumberREBU1047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2018
Event Location Hospital
Initial Date Received by Manufacturer 02/01/2018
Supplement Date Received by Manufacturer03/08/2018
Initial Report FDA Received Date02/06/2018
Supplement Report FDA Received Date04/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-