Model Number ZXT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Halo (2227); No Code Available (3191)
Date of Event 06/27/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt150 22.5 diopter intraocular lens (iol) was implanted into the left eye (os) of the patient on (b)(6) 2017.It was later explanted on (b)(6) 2018, due to halos and glare.There was no incision enlargement, vitrectomy, or sutures.The lens was replaced with zcboo 22.5 diopter iol.Reportedly, there was no patient injury and the patient is doing good post-operatively.No further information was provided.
 
Additional Manufacturer Narrative
The product testing was not performed as the complaint device was not returned for evaluation.The reported complaint cannot be verified.Manufacturing records were reviewed and the device was manufactured according to specifications.A complaint history search revealed no similar complaints were received for this production order number to date.The directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7246328
Report Number9614546-2018-00118
Device Sequence Number854080
Product Code POE
UDI-Device Identifier05050474605251
UDI-Public(01)05050474605251(17)220126
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date01/26/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U225
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/10/2018
Supplement Date Received by Manufacturer06/06/2018
Initial Report FDA Received Date02/06/2018
Supplement Report FDA Received Date06/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-