• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Device Problems Calibration Error (1078); False Reading From Device Non-Compliance (1228); Device Displays Incorrect Message (2591)
Patient Problem Hyperglycemia (1905)
Event Date 01/27/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm, experienced high blood glucose level and change sensor alert.The customer¿s blood glucose was 260 mg/dl and the sensor glucose was 65 mg/dl at the time of the incident.Customer received a change sensor alert after a calibration error.Troubleshoot was performed and the customer stated that insulin delivery was suspended low sensor glucose.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.Customer declined to troubleshoot for high readings.The sensor will not be returned for analysis.
 
Manufacturer Narrative
Inspected one opened and used sensor.We performed continuity resistance test and sensor passed per specifications.Performed bicarbonate buffer test sensor passed per specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7246365
MDR Text Key99181125
Report Number3004209178-2018-48768
Device Sequence Number1
Product Code OZO
UDI-Device Identifier20643169786725
UDI-Public(01)20643169786725(017)20180517
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date05/17/2018
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG253ZS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/27/2018
Initial Date FDA Received02/06/2018
Supplement Dates Manufacturer Received05/29/2018
Supplement Dates FDA Received06/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age55 YR
Patient Weight190
-
-