• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX OTW; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX OTW; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX27515W
Medical Device Problem Codes Device Damaged by Another Device (2915); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The physician intended to use a resolute onyx drug eluting stent to treat a mildly tortuous, mildly calcified lesion in the mid lad.No abnormalities were reported in relation to anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop.The device was inspected with no issues.Negative prep was performed with no issues.The lesion was pre-dilated.The device passed through a previously deployed stent.No resistance was encountered when advancing the device.Excessive force was not used.The physician was trying to pass the resolute onyx stent through existing older unknown brand stent, and it got caught on the previously placed stent.The device failed to cross the lesion.It was reported that the stent deformation occurred in vivo during positioning.The physician completed the procedure with a medtronic product.No patient injury was reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX OTW
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7246909
Report Number9612164-2018-00216
Device Sequence Number13690433
Product Code NIQ
UDI-Device Identifier00643169602410
UDI-Public00643169602410
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2019
Device Catalogue NumberRONYX27515W
Device Lot Number0008493579
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/18/2018
Supplement Date Received by Manufacturer10/11/2018
Initial Report FDA Received Date02/07/2018
Supplement Report FDA Received Date10/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age51 YR
Patient Weight98
-
-