C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE Back to Search Results
Model Number 0102L18
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter balloon deflated on it's own.No missing pieces were reported and no medical intervention was required.
 
Additional Manufacturer Narrative
Received 1 used red lubricath.The reported event was unconfirmed since the problem could not be reproduced.The exterior of the sample was examined and did not find anything to contribute to the reported event.The balloon was inflated with air while submerged in water and the balloon did not leak.A potential root cause is a pinhole in the balloon.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.".
 
Event or Problem Description
It was reported that the catheter balloon deflated on it's own.No missing pieces were reported and no medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE
Common Device Name
LUBRICATH
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7247160
Report Number1018233-2018-00363
Device Sequence Number16383072
Product Code EZC
UDI-Device Identifier00801741018138
UDI-Public(01)00801741018138
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/22/2020
Device Model Number0102L18
Device Catalogue Number0102L18
Device Lot NumberSD5017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Initial Date Received by Manufacturer 01/18/2018
Supplement Date Received by Manufacturer04/05/2018
Initial Report FDA Received Date02/07/2018
Supplement Report FDA Received Date04/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-