ANGIODYNAMICS STARBURST¿ TALON ELECTRODE; ELECTROSURGICAL PROBE, Back to Search Results
Catalog Number H7877001028471
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.The firm is attempting to obtain the device.An investigation into the root cause for product problem is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
As reported to angiodynamics on january 23, 2018: during preparation for an rfa procedure, one of the needles (tine) of the talon probe detached.The device was set aside and a new of the same was used to complete the procedure.It was reported that the device involved in the incident is available to be returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was a starburst talon probe.A visual review of the returned device noted the needles were deployed and array #2was damaged and array #3 was fractured.The window of array #3 was viewed under a microscope and a the fractured off piece appears to stuck in between the drive tube and the deployment window.A functional test was performed.The tines were able to be deployed and retracted repeatedly without issue.The customer's reported complaint description of a tine breaking off is confirmed.The fractured piece of tine was visualized stuck between the drive tube and array deployment window.This event occurred outside the body and the patient was unaffected.This device is fully deployed in the device tray when shipped from the manufacturing facility.Though it cannot be definitively determined, the most likely root cause to this event is due to handling during preparation or while removing the device from the packaging.If a tine becomes bent/damaged while deployed, it's possible for the damaged tine to become trapped in the window during subsequent deployment and causing it to break under force.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instruction for use, which is supplied to the user with this catalog number, contains the following statements: inspect the device for any damage.Do not use a device that has been damaged.Do not twist or exert high forces on the device while it is deployed in the tissue.This may cause the needles to break and remain in the tissue.Do not remove the device without ensuring that the needles are fully retracted within the trocar.This may cause the needles to break and remain in the tissue.If retraction of the device becomes difficult, do not exert additional force.Infuse a small amount of saline solution and gently work the needles loose by alternating between a slightly retracted and a deployed position (while holding the main body with one hand and the trocar (at the point of insertion) with the other hand).The saline solution will generally soften the ablated tissue surrounding the electrodes.Once the ablated tissue softens, the arrays can be worked loose from the tissue and fully retracted." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint reference # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARBURST¿ TALON ELECTRODE
Common Device Name
ELECTROSURGICAL PROBE,
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key7247628
Report Number1056436-2018-00003
Device Sequence Number18859622
Product Code GEI
UDI-Device IdentifierH7877001028471
UDI-PublicH7877001028471
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K010060
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/31/2020
Device Catalogue NumberH7877001028471
Device Lot Number5232990
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2018
Supplement Date Received by Manufacturer01/23/2018
Initial Report FDA Received Date02/07/2018
Supplement Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-