• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012455-20
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Evaluation summary: visual and dimensional inspections were performed on the returned device.The reported separation was confirmed; however, the difficulty removing the device from the guiding catheter could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported complaints appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that a 5.0x20mm nc trek advanced to the proximal right coronary artery lesion and post-dilatation was performed without issue on an unspecified stent.Following balloon deflation without issue, the device was removed into the guide catheter.Once inside the guide catheter, resistance was met and the nc trek dilatation catheter and the guide catheter were removed as a single unit.Once outside the patients anatomy, the nc trek shaft was pulled and shaft separated.There were no adverse patient effects and there was no clinically significant delay.There was no additional information provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7247968
Report Number2024168-2018-00881
Device Sequence Number1031888
Product Code LOX
UDI-Device Identifier08717648152160
UDI-Public08717648152160
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Catalogue Number1012455-20
Device Lot Number61028G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2018
Initial Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
-
-