Model Number S7 |
Device Problems
No Display/Image (1183); Device Stops Intermittently (1599)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was unavailable from the site.A medtronic representative went to the site to test the equipment.Troubleshooting found that the power supply for the navigation system required replacement.The representative then returned to the site and replaced the power supply to resolve the reported issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
A site representative reported that, while in a procedure, the surgeon monitor lost display functionality.There was no known impact on patient outcome.No additional information was provided.
|
|
Manufacturer Narrative
|
The power supply for the navigation system was returned to the manufacturer for evaluation.Testing found that the chassis had an open fuse on the board.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
|
Manufacturer Narrative
|
Additional information: additional information was received from a manufacturer representative.It was reported the procedure proceeded with the staff cart monitor.There was no negative impact on the patient.
|
|