Model Number ZXR00
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Codes Blurred Vision (2137); Vitrectomy (2643); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a tecnis symfony multifocal intraocular lens (model zxr00 +24.0 diopter) was explanted in a secondary procedure because of patient complaining of unsatisfactory visual outcome, hazy vision, and doctor's observation of a displaced haptic.During the explant procedure doctor had to perform a vitrectomy and an non-amo intraocular lens with +22.0 diopter was implanted as the replacement lens for the patient.There was no patient injury post op.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 02/21/2018.Device returned to manufacturer? yes.Device evaluation: the sample was returned to the manufacturer.Visual inspection was performed and product was cut in half, most probably to make the explant possible.Considering the condition of the returned lens, further analysis is not possible.The reported complaint cannot be confirmed.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7248871
Report Number9614546-2018-00120
Device Sequence Number834660
Product Code POE
UDI-Device Identifier05050474579170
UDI-Public(01)05050474579170(17)220409
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/09/2022
Device Model NumberZXR00
Device Catalogue NumberZXR00U0240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2018
Initial Date Received by Manufacturer 01/10/2018
Supplement Date Received by Manufacturer04/06/2018
Initial Report FDA Received Date02/07/2018
Supplement Report FDA Received Date04/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1