| Catalog Number |
8267.VIT23 |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/22/2014
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Incident happened during surgery, bubbles popping out from cutters hence concluded as reportable.The risk is injury by microbial, viral or prion infection (patient harm).There is no information that reasonably suggests a patient injury occurred.Investigation could not reveal the root cause, due to not receiving the faulty article.Therefore this complaint is closed.This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic research centre (b)(4)).Details of this activity were discussed with cdrh office of compliance (b)(6) during a tele-conference on 05/31/2017.All available information has been provided.
|
| |
|
Event or Problem Description
|
|
A report was received of an incident occurring in (b)(6).Previously,we also noticed the bubbles popping out from cutter.We simply change the cutter during the surgery.No further information is available.
|
| |
|